OMNIVERSA
®
MULTI-MODALITY THERAPY USER MANUAL
- 101 -
COPYRIGHT © 2017-2020, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
TECHNICAL SPECIFICATIONS (cont.)
PENS (PATTERNED ELECTRICAL NEUROMUSCULAR STIMULATION) MODE:
Output/Waveform:
0 – 100 mA peak, 500 Ω load; Asymmetric Biphasic – Based on EMG functional pattern
Phase Duration:
Variable from 40 µs to 100 µs
Pulse Rate:
Set at 50 Hz or 100 Hz
HVPC (HIGH VOLTAGE PULSED CURRENT) MODE:
Output/Waveform:
0 – 225V; Twin monophasic pulses each with 40 µs phase duration
Phase Duration:
< 2 uSec
Pulse Rate:
Adjustable 0 – 200 Hz
Interphase Interval:
Preset at 65 µs
HAND CONTROL:
MFAC Hand Control:
Modified Sine wave into physiologic load 2 KHz, 2.5 KHz,3 KHz, 3.3 KHz , 4 KHz , 5 KHz ,
6 KHz , 7 KHz , 8 KHz , 9 KHz ,and 10 KHz carrier frequency.
Burst rate adjustable from 0.1 to 100bps, 50% duty factor
Rate scan from 0-50% programmable with scan time, programmable from 0-20 seconds
LVPC Hand Control:
Asymmetrical biphasic square wave
Phase duration adjustable from 10 - 300 µs
Pulse rate adjustable from 0.1 – 200 pps
TIMER FUNCTIONS:
Treatment Timer:
Adjustable for 0-60 minutes in one minute increments.
Channel Timing:
In alternating mode channels follow each other sequentially. In simultaneous modes the output
is simultaneous. In delayed mode channel B is delayed.
TIMING SELECTIONS:
Ramp Time:
Adjustable from 0 to 9 seconds in 1-second increments. Ramps overlap in NMES Alt mode or
with channel delay.
ON Time:
Adjustable from 1 to 60 seconds in 1-second increments
OFF Time:
Adjustable from 0 (continuous) to 120 seconds in 1-second increments
BATTERIES:
Type:
Gel Cell rechargeable 12V /2AH
Charge Time:
4-hour charge time is required for a full charge
MISC.
Useful Life
The device is deemed to have a useful life of 5 to 7 years. This is largely dependent on usage
and maintenance (including cleaning after each use).
CAUTION:
Federal law restricts this device to sale by or on the order of a physician (or other health practitioner
licensed by their State).
ACP reserves the right to change technical specifications and product availability without notice.
(EMC) Electromagnetic Compliance
Medical electrical devices series are subject to special precautions with regard to electromagnetic compatibility and must be
installed and commissioned in accordance with the advice given in the instructions for use and accompanying documents.
Portable and mobile RF communication systems (e.g. mobile phones) may interfere with medical devices. The OmniVersa