Nautilus Data Loggers
11
Copyright © 2010 ACR Systems Inc. All Rights Reserved.
the Realtime value using the currently selected equation is shown. *
* Note that if a custom equation is not available when Edit Setup... is selected (e.g.: deleted)
then the equation will be shown as unknown. If a new equation is not selected the Realtime
value will not display correctly. If the logger is backed up the channel will be disabled then
changes to the equation can be made in the graph using the Lines Tab in Graph Settings."
1.2.2.2
Channel Calibration
The Calibration subsection shows the Low, Mid, High, and Ext calibration values. These
values are used to adjust the readings of the logger in order to make the output more
accurate. To set each calibration value, click on the appropriate calibration field and enter the
value. The factory equations for the Nautilus use the Low, Mid, and High values. Ext is not
used.
Note that the calibration values are based on 255 increments between -5 and +5, the field will
default to the closest value of what is entered. E.g.: -4.9 will default to -4.8828.