APPENDIX A  TRH-1000 Sample Rates

TRH-1000 Sample Rates 

1.4.1

How a Data Logger Stores Readings

The sample rate of your data logger is the frequency with which it stores readings in its
memory.  A selected sample will apply for all channels on a TRH-1000.

If you set your data logger's sample rate to eight seconds, it reads its input channels and
stores the readings once every eight seconds.  If you choose a sample rate longer than eight
seconds, the logger still reads its input channels once every eight seconds, and stores an
averaged reading at the end of the sample interval you have chosen if the sampling mode is
set to Average.

For example, if you use a two-minute sample rate, the data logger takes fifteen separate
readings over each two-minute interval.  At the end of each two-minute interval, the logger
calculates the average of the fifteen readings, and stores that average in its memory.

Otherwise if the sampling mode is set to Spot the logger still takes readings every 8 seconds
but only stores the reading taken at the selected interval.

1.4.2

Sample Rate Table

The following tables list the maximum time spans over which your TRH-1000 will record
readings.  The time spans depend on the length of the sample rate.  A TRH-1000 has a
memory capacity of 32,640 readings per channel.

In each table, the Sample Rate column lists the available sample rates.  The Days, Hours,
Minutes
 and Seconds columns list the time spans over which your logger will record readings
at each sample rate.

For example, a TRH-1000 using a sample rate of ten minutes will store readings for 226 days
and 16 hours.  After that time, depending on how it was set up, the logger will either continue
to take readings and replace the oldest readings with the new readings (FIFO), or it will stop
taking readings (Stop When Full).

Table B-1:  TRH-1000 Sample Rate Table

Sample Rate

Years

Days

Hours

Minutes

Seconds

 8 sec.

0

3

0

32

0

 16 sec.

0

6

1

4

0

www.

GlobalTestSupply

.com

Find Quality Products Online at:

sales@GlobalTestSupply.com

Summary of Contents for 01-0194

Page 1: ...Copyright 2010 ACR Systems Inc All Rights Reserved Reference Guide IMPORTANT TRH Data Loggers www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Page 2: ...For more information regarding trademarks see the Trademark Notices section of this manual ACR Systems Inc assumes no responsibility for errors or omissions or for damages resulting from the use of i...

Page 3: ...and Canadian Customs clearance if applicable Any replacement hardware or software will be warranted for the remainder of the original warranty period To the maximum extent permitted by applicable law...

Page 4: ...Approvals 7 2 TRH Software Driver 7 Logger Status 7 Logger Sampling 8 Logger Memory 10 Logger Channels 10 Channel Information 11 Channel Calibration 12 3 TRH 1000 Temperature Relative Humidity Logger...

Page 5: ...conventions and procedures To get the most from your logger please take time to familiarize yourself with TrendReader Standard and its Reference Guide For important information on how TRH 1000 data lo...

Page 6: ...e them a useful tool for data collection Pocket sized Solid state components no moving parts Low power microprocessor controlled circuitry 8 bit resolution readings High capacity lithium battery Quart...

Page 7: ...ing one of two sampling methods 1 First in First out FIFO In this method the logger continues to take readings when its memory is full to make room it discards the oldest reading every time it adds a...

Page 8: ...puter You can then configure your logger with various options by talking to it via the interface cable Set Sample Rate Always confirm the sample rate how often the logger saves readings to make sure i...

Page 9: ...using a protective enclosure For humid conditions you can protect your logger simply by placing it in a zip lock plastic bag Note that covering the logger may affect the relative humidity readings Wh...

Page 10: ...ge 1 Continuous First in First out 2 Stop when full Fill then stop Sample Intervals User selectable rates from every 8 seconds to 34 minutes Sampling Mode 1 Average over sample interval except for 8 s...

Page 11: ...software Specifications are subject to change without notice 1 1 7 Product Approvals Certified to CE standard EN61326 1977 A1 1998 European Emissions and Immunity covering ESD RFI EFT B Surge Conducte...

Page 12: ...with the existing or new parameters select Edit Setup from TrendReader In the Edit Logger Setup window make any necessary changes To implement your changes you can either click on Apply or OK OK will...

Page 13: ...nd what the start and end times or capacity are for the data logging In the FIFO First In First Out mode the logger continually stores data and when the memory limit is reached the new readings start...

Page 14: ...op up a form allowing you to set the date and time Highlight the date or time and enter the appropriate value You can also click on the drop down list arrow to bring up a calendar You can set the logg...

Page 15: ...your own custom equations See Equations in the TrendReader guide for further information To set the equation click on the Equation field and select the equation from the drop down list To help you se...

Page 16: ...ccurate To set each calibration value click on the appropriate calibration field and enter the value The factory equations for the TRH 1000 use the Low and Mid values as offset and span High and Ext a...

Page 17: ...sories are required making the TRH 1000 the ideal data logger for quick and accurate temperature and humidity measurements in a variety of applications 1 3 2 Setup Once the logger is connected to your...

Page 18: ...ly from location to location Such factors as air stratification drafts and proximity to heat or humidity sources people equipment moisture solar gain etc can contribute to a wide variance in condition...

Page 19: ...d stores that average in its memory Otherwise if the sampling mode is set to Spot the logger still takes readings every 8 seconds but only stores the reading taken at the selected interval 1 4 2 Sampl...

Page 20: ...Inc All Rights Reserved 32 sec 0 12 2 8 0 56 sec 0 21 3 44 0 2 min 0 45 8 0 0 4 min 56 sec 0 111 19 44 0 10 min 0 226 16 0 0 20 min 1 87 36 0 0 30 min 1 315 0 0 0 www GlobalTestSupply com Find Quality...

Page 21: ...el Information 10 Channels 10 Description 1 Features 1 How to Use 3 Memory 8 Operation 2 Readings 2 Sampling 7 Setup 7 Software 7 Specifications 5 Status 7 Unknown Equation 10 TRH 1000 12 Approvals 6...

Reviews: