Click on the
pull-down menu and select
Cal Stat. Read
. A Transaction Screen similar to
the following will be displayed:
At this stage, proceed to the CS and complete the following procedure.
1. Connect to ECU on CS.
2. Enter with the USER ID for the as follows:
a) Select
Diagnostic
(number 3 on key pad)
b) Select
Previous
(number 6) - the ECU screen will show
Enter ID
c) Select
Odometer
(Button #) on the genius to enter # plus 3 digits
d) Press
OK
(number 7)
3. The ECU will respond with a double-beep to confirm the number was entered correctly and the screen will show
OK
.
NOTE:
Leave the calibration area after connecting the Genius to the PC cable.
4. Return to ITE and press the
Save and Print
button to start the transaction.
The
Monitor
screen will appear, and the progress of the read transaction can be observed.
5. The ECU will sound a triple-beep to indicate that the transaction is complete.
The CS is now ready for calibrating the HS and recording activity for the next readout.
12.2
Result
The contents of the Auto-Cal ECU are transferred to the Service Centre PC for transfer to the central computer during
synchronization. There is no printout of log details to the Service Centre, but data management staff may review the record to
discern any departures from procedures. Some international jurisdictions require these records for ongoing certification.
13.0
Reusing mouthpiece connectors
The condensation may be removed from the mouthpiece connector by connecting it to the aquarium air pump when
“foggy” and
before condensation pools into fluid.