19
Avoiding contamination
In all modes except unmodulated light, the aver-
age power drops by 3 dB. This is due to the mod-
ulation on which the signal coding of the AU
mode is also based.
Avoiding contamination
The FC-PC connector system uses direct con-
tact between the optical fibers. Even the slightest
contamination such as dust particles or sweat
from your hands (fingerprints) can lower the
measurement accuracy (insertion loss, reflec-
tion). This can also cause permanent damage
such as scratches on the optical surfaces. This
applies to the connectors on your OLS-6 and
also to the connectors on the test cables.
Tip:
Þ
Clean the connectors on your test cables be-
fore connecting them by wiping off the sur-
face with a piece of cleaning strip.
Þ
When you take a break, always close the
protective cap if you remove the test cable
from the connector.
Þ
Clean at regular intervals.
If over time the output level deviates from the
normal value, the test connector might be soiled.
For information on cleaning, see page 25.