background image

 

1

 

 

 

ActiveCare

    TD-4121 

Blood Glucose 

Multi-Functional Monitoring System 

 

Owner

’s Manual 

Summary of Contents for TD-4121

Page 1: ...1 ActiveCare TD 4121 Blood Glucose Multi Functional Monitoring System Owner s Manual...

Page 2: ...2...

Page 3: ...using this product please read the following instructions thoroughly and attentively Regular monitoring of your blood glucose levels can help you and your doctor gain better control of your diabetes Y...

Page 4: ...anufacturer 3 Do NOT use the device if it is not working properly or if it is damaged 4 Do NOT under any circumstances use the device on neonates or infants 5 This device does NOT serve as a cure for...

Page 5: ...tial to the durability of your device If you are concerned about its accuracy please contact TaiDoc or place of purchase for help Action upon Delivery Unpacking checking delivery for completeness chec...

Page 6: ...ing 23 Performing a Control Solution Test 23 TESTING WITH BLOOD SAMPLE 25 Preparing the Lancing Device for Blood Testing 25 Preparing the Puncture Site 25 Performing a Blood Glucose Test 29 Reviewing...

Page 7: ...are experiencing symptoms that are inconsistent with your blood glucose test results and you have followed all the instructions given in this owner s manual contact your healthcare professional We do...

Page 8: ...used by a single person and should not be shared It should not be used for the diagnosis of or screening for diabetes or testing on neonates Professionals may test with capillary and venous blood samp...

Page 9: ...2 Owner s Manual 3 Quick Start User Guide 4 Test Strips 5 Protective wallet 6 1 x 3 7V Li Polymer rechargeable battery 7 Lancet NOTE If any items are missing from your kit or opened prior to use pleas...

Page 10: ...charge 5 Home Button To enter the meter function and click for confirmation or exit 6 Display Window and Touchscreen The touchscreen allow users to interact with what is displayed directly by the fing...

Page 11: ...g which are shown on the main page Other icons on the Main Page are introduced as below 1 Battery 2 Alarm Icon 3 Time 4 Flight Icon 5 GPRS Signal 6 Date 7 Measuring 8 Memory 9 Setting 10 Calling Indic...

Page 12: ...nt Hole Apply a drop of blood here The blood will be automatically absorbed Confirmation Window This is where you confirm if enough blood has been applied to the absorbent hole in the test strip Conta...

Page 13: ...he system off Press the power button for 1 second to activate the meter The meter displays ActiveCare logo with a loading bar showing the process of time date synchrony You will see the main page show...

Page 14: ...ferred mode airplane or general for preference Select the airplane mode when necessary or general mode to activate the meter s telecommunication function 3 Tap on to save and exit Note The airplane mo...

Page 15: ...set the measuring unit There are two measuring units to choose from mg dL and mmol L 2 Tape on the desired one and to save and exit Volume 1 Tape to set up the volume 2 Tape on the or to adjust the vo...

Page 16: ...mode In this mode time and date are synchronized once the meter s GPRS General Packet Radio Service is in service Please contact your mobile communication providers for further information on GPRS 3 T...

Page 17: ...ime by tapping on or In 12 hour time setting you may choose from AM and PM along with time adjustment 4 Tap on to save and exit 5 Press the home button to leave the alarm setting page and back to the...

Page 18: ...the displayed language The default language is English 2 Tap on the preferred language bar 3 Tap to save and exit Congratulations You have completed all settings NOTE These parameters can ONLY be cha...

Page 19: ...ement To transmit the test result to the server The service center will phone in if the uploaded result is atypical 2 At 6 o clock every morning To transmit data to the server automatically providing...

Page 20: ...the call Appears a call comes in Tap on to take or to reject the call Appears when you are answering the phone Tap on or to adjust the volume Tap on to hang up the phone Appears when the meter is sea...

Page 21: ...for immediate assistance The phone number of the service center is programmed prior to your receipt of the meter If not please contact the customer service for assistance Please contact your service p...

Page 22: ...halt state 3 At the halt state press the home button Then keep on pressing the home button with another finger pressing the power button for 1 second Then release both fingers at the same time 4 Tap...

Page 23: ...sting process You have dropped or think you may have damaged the meter Performing a Control Solution Test 1 Start with the meter off Press the power button for 1 second to turn on the meter Wait for t...

Page 24: ...n test result will appear on the display in seconds Compare this result with the range printed on the test strip vial and it should fall within this range If not please read the instructions again and...

Page 25: ...meter testing the assistant should wear gloves throughout Additionally the assistant should clean and disinfect meter between each use Preparing the Lancing Device for Blood Testing Please follow the...

Page 26: ...than the fingertip Replace the lancing device cap with the clear cap for AST Pull the cocking control back until it clicks When lancing the palm the forearm and the upper arm you must avoid lancing t...

Page 27: ...he body other than the fingertips to check their blood glucose levels This system allows you to test on the palm the forearm or the upper arm with results equivalent to fingertip testing What is the a...

Page 28: ...n 2 hours since the last meal Two hours or more after taking insulin Two hours or more after exercise Do NOT use AST if You think your blood glucose is low You are unaware of hypoglycemia You are test...

Page 29: ...o the strip slot to start measuring When the main page appears tap on to initiate the blood glucose measurement procedure 2 Perform the blood glucose test as directed on the display Insert a test stri...

Page 30: ...ive inaccurate glucose results 5 Wait a few seconds for the measurement to process When the page of measuring mode appears tap on or to choose the mode that fits you the best General or after before b...

Page 31: ...TE Do not press the punctured site against the test strip or try to smear the blood If you do not apply a blood sample to the test strip within 10 minutes the meter will automatically turn into the ha...

Page 32: ...dates and times in the memory feature If the memory bank is full the latest result will overwrite the oldest one 1 Turn on the Meter Start with the meter off Press and release the power button to act...

Page 33: ...he meter into a wall outlet to initiate the charging process The battery charging indication will light red when charging is in process When the indicator light green it means the battery is full You...

Page 34: ...The used meter should be treated as contaminated that may carry a risk of infection during measurement The batteries in this used meter should be removed and the meter should be disposed in accordanc...

Page 35: ...urate results Do not bend cut or alter a test strip in any way Keep the strip vial away from children since the cap and the test strip may be a choking hazard If swallowed promptly see a doctor for he...

Page 36: ...a pure sample and an accurate result Store the control solution tightly closed at temperatures between 2 C to 30 C 35 6 F to 86 F Do NOT freeze NOTE Do NOT clean and disinfect the meter while perform...

Page 37: ...ted Repeat the test with a new test strip Contact the service center if the problem persists Timeout The procedure has taken too long Sample not enough Appears when the amount of blood taken is not en...

Page 38: ...tting No Strip detect Appears when the strip is not inserted correctly Please re insert the strip and make sure its front side is facing up TEMP too low Appears when ambient temperature is above or be...

Page 39: ...new test strip Defective meter Please contact customer service 3 If the control solution testing result is out of range POSSIBLE CAUSE WHAT TO DO Error in performing the test Read instructions thoroug...

Page 40: ...the battery Power Source one Li Polymer rechargeable battery Display LCD Memory 450 measurement results with respective date and time External Output General Packet Radio Service GPRS Auto electrode i...

Page 41: ...nce of 0 5 cm must be maintained between the user s body and the device including the antenna during body worn operation to comply with the RF exposure requirements in Europe To compliance with RF Exp...

Page 42: ...cal device Caution consult accompanying documents Consult instructions for use Humidity Limitation Temperature limitation Collection for electrical and electronic equipment Use by CE mark Batch code M...

Page 43: ...this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference b...

Page 44: ...ement known as the Specific Absorption Rate or SAR The SAR limit set by the FCC is 1 6W kg Tests for SAR are conducted using standard operating positions accepted by the FCC with the device transmitti...

Page 45: ...45 TaiDoc Technology Corporation B1 7F No 127 Wugong 2nd Rd Wugu Dist 24888 New Taipei City Taiwan www taidoc com For self testing MedNet GmbH Borkstra e 10 48163 M nster Germany...

Reviews: