50
I
II - 1 Installation Set
u
p
Starting a Non-Linear Error Compensation Table
Select Non-linear by pressing the TYPE soft key.
To start a new error compensation table, first press the EDIT TABLE
soft key.
All correction points (up to 200) are equally spaced from the start
point. Enter the distance between each of the correction points.
Press the DOWN arrow key.
Enter the table’s start point. The start point is measured from the
scale’s reference point. If this distance is not known, you can move
to the location of the start point, and press TEACH POSITION.
Press ENTER.
Configuring the Compensation Table
Press the EDIT TABLE soft key to view the table entries.
Use the UP, or DOWN arrow keys, or the numeric keys to move the
cursor to the correction point to be added or changed.
Press ENTER.
Enter the known error which exists at this point. Press ENTER.
When completed, press C key to exit the table and return to the
Error Compensation form.
Reading the Graph
The error compensation table may be viewed in table or graphical
formats. The graph shows a plot of a translation error vs. measured
value. The graph has a fixed scale. As the cursor is moved through the
form, the location of the point on the graph is indicated with a
vertical line.
Pressing ENTER will save the spacing, and start
point information.
Summary of Contents for 200S READOUTS
Page 1: ...200S READOUTS REFERENCE MANUAL ...
Page 2: ......
Page 6: ...iv ...
Page 8: ...vi ...
Page 10: ...viii ...
Page 15: ...1 Operating Instructions ...
Page 58: ...44 I I 4 Turning Specific Operations ...
Page 59: ...45 Technical Information ...
Page 78: ...64 I II 6 Dimensions ...
Page 81: ......