POWER

 

SUPPLY 

 
Power  for  the  relays  can  come  from  a  user  supplied  source.    2 
Alternative power input points are provided depending upon the power 
the user has available. 
 
Connector  VAC  is  used  for  15-25VDC  or  12-20VAC,  50/60Hz.    500ma. 
This input is not sensitive to polarity, either terminal can be used. 
 
If the user has 12VDC to 15VDC, 500ma available then connector marked 
12VDc  can be used.   Be careful with the  polarity, using +12  to 12V and 
minus voltage to the GND terminal.  
 
Only one power input terminal should be used at one time 
 

OPERATION 

 
During  operation  the  relays  will  be  active  at  the  same  time  the 
Low/OK/High  LEDs  on  the  front  panel  operate  and  the  Zero  relay  is 
activated when the scale is at zero point. 
 
Note  the  scale  must  show  a  small  weight  before  the  LOW  LED  will 
activate.   See the AE402 User Manual for details of how to set the limits. 
 
User  connections  to  the  relays  are  made  using  the  terminal  strips 
marked OUT1 to OUT4.  The relay active for each LED will depend on the 
input connection, Input IN1 controls relay 1, OUT1 for example.  But IN1 
could be connected to any of the 4 terminals in the AE 402. 
 

CAUTION: 

The voltage and current limits of the relay outputs must be less than: 
115/230VAC, 1amp  or   24VDC. 1amp. 
 
Higher currents may damage the circuit board. 
The relays are not specified for higher voltages. 