12 ERROR CHECKING
During weighing the balance is constantly checking to see if it is operating within the
limited parameters. The errors likely to occur are:
A/D counts below lowest allowed value
A/D counts above highest allowed value
A/D not operating
Maximum capacity exceeded
Other errors may be detected during special functions or operations. These will be
described in the section that applies.
Error messages and the reasons are:
Concerning A/D counts
ERR UL
ERR UL
ERR UL
ERR UL
A/D counts below a limit
ERR OL
ERR OL
ERR OL
ERR OL
A/D counts above a pre-set limit
Concerning calibration
ERR STB
ERR STB
ERR STB
ERR STB
Calibration could not be completed
because the results were not stable
ERR LO
ERR LO
ERR LO
ERR LO
or
ERR HI
ERR HI
ERR HI
ERR HI
Calibration constant not within 20% of old
calibration constant
Concerning weighing
ERR LO
ERR LO
ERR LO
ERR LO
Weight display is below zero by >4%max
ERR HI
ERR HI
ERR HI
ERR HI
Weight is above maximum plus 90d