6.0 CFW USER CALIBRATION
The scale can be calibrated using the following procedure. For a more
detailed method of calibrating the scale, it will be necessary to enter the
secure Programming Menu. Refer to the CFW Service Manual for more
information.
CFWa Scales only:
CFWa
scales will also have the lb, oz, lb-oz or kg (or g) indicator on to show the unit
of the weight requested. If the scale was in pounds before starting the calibration,
the weights requested will be in pound values. If the scale was weighing in kilograms
then metric weights will be requested.
PROCEDURE
•
Turn the power off.
•
Turn the power back on, during the counting from 1 to 0, press
[Func/C]
.
•
The display will show "
CAL
". While it is showing "
CAL
" press the
[Unit]
,
[Print]
and
[Tare]
keys in sequence to enter the Calibration
section. The display will show "
unLoAd
".
•
Remove any weight from platform when stable and press
[Zero/Enter]
.
•
The display will show the last calibration weight used. If this is
correct you can continue by pressing
[Zero/Enter]
. If it is not
correct use the arrow keys to change the calibration weight value.
When it is correct press
[Zero/Enter]
.
•
The display will show "
LoAd
". Place the calibration weight on the
scale. Press the
[Zero/Enter]
key.
•
If the calibration is acceptable the scale will run a self-test during
which the calibration weight should be removed. If an error
message “
FAiL L
” is shown try calibration again as a disturbance
may have prevented a successful calibration.
@Adam Equipment Company 2007
11