2.0 TROUBLE SHOOTING GUIDE
PROBLEM
POSSIBLE CAUSE
Display is blank
Power Switch is faulty or not turned on
No turn on test
Battery is faulty or not charged
Power supply is incorrect
Display all zeros after turn on
Pan not installed
or E6 is displayed
Unstable weight
Load cell damaged
Electronics
faulty
‘- - - - -‘ appears on display
Maximum capacity exceeded
Load Cell damaged
Electronics is faulty
Display is unstable
Drafts or air currents
Obstruction under pan
Sample is moving (animal weighing)
Vibrations through table
Temperature changed dramatically
Electronics
faulty
Battery
low
Weight value incorrect
Calibration error, Recalibrate
Unit calibrated with inaccurate weight
Balance not level
Obstruction between sample and cover
Wrong unit of weight displayed
Cannot use Full Capacity
Over-load Stops hitting pan support or hitting
bottom
of
load
cell
Electronics
faulty
Parameters
set
incorrectly
Load
Cell
Damaged
Not Linear
Overload stops hitting too soon
Faulty
Factory
Calibration
Off Centre Loading error
Overload Stops not correct
Obstruction
under
pan
Battery will not charge
Charging circuit failure
Battery
failure
Incorrect
adapter
(below
800
mA)
Processor failure
@Adam Equipment Company 2007
5