12704A Resistivity Chamber Operation Manual
3.2 Setting the Electrodes and a Sample
3.2 Setting the Electrodes and a Sample
Use a sample that meets the following conditions. Measure the thickness of the
sample previously with a vernier caliper or micrometer.
85 mm diameter or more
150 (W) × 5 (H) × 110 (D) mm or less
(1)
Open the cover and check that the conductive rubber main electrode and ring electrode do
not stick out of the plates respectively.
(2)
Select volume resistivity measurement or surface resistivity measurement by the RESIS-
TIVITY switch.
Volume resistivity measurement: Set the RESISTIVITY switch to VOLUME.
Surface resistivity measurement: Set the RESISTIVITY switch to SURFACE.
(3)
Set the PROTECTOR switch.
ON: If a sample is short-circuited due to insulation breakdown, the electrodes will be
damaged. In such a case, turn the switch to ON.
1 M
is added to measured values. It takes longer time to reach a true value after
measurement starts.
OFF: If there is no possibility that a sample is short-circuited, turn the switch to OFF.
When the current limit of the V SOURCE is 10 mA or less and the generated volt-
age is 10 V or less, measurement can be done with the switch set to OFF even if
there is a possibility that the sample is short-circuited.
(4)
Set a sample onto the main electrode and the guard electrode so as to cover them com-
pletely.
(5)
Pass the feed terminal through the two holes on the opposed electrode, and press the op-
posed electrode down to the end to stick fast to the sample.
(6)
Close and lock the cover.
Open the cover and make sure that the applied voltage switch is OFF or voltage is
not being applied when operating the RESISTIVITY switch. Operating the switch
with voltage being applied will cause damage to the switch.
CAUTION
WARNING
Summary of Contents for 12704A
Page 2: ......
Page 4: ......
Page 5: ...1 1 12704A Resistivity Chamber Operation Manual 1 Introduction Chapter 1 Introduction ...
Page 14: ......
Page 23: ...4 1 12704A Resistivity Chamber Operation Manual 4 Specifications Chapter 4 Specifications ...
Page 25: ...A 1 12704A Resistivity Chamber Operation Manual APPENDIX APPENDIX ...