4601 I-V Meter Operation Manual
7. PERFORMANCE TEST
7-1
7.
PERFORMANCE TEST
This chapter describes the methods for checking whether the 4601 operates in the specified accuracy.
7.1
Measuring Instruments Necessary for Performance Test
Use measuring instruments for the performance test which meet the specifications described in Section 8.1.1,
"Measuring Instruments Necessary for Calibration."
7.2
Connections
The connections required for the performance test are the same as shown in Figure 8-4, "Connections for Cal-
ibration(4 of 4)."
7.3
Test Method
Execute the performance test under the following conditions in a location free of dust, vibration, noise or
other adverse conditions:
Temperature:
23
C
5
C
Relative humidity: 60% or lower
Warm-up:
60 minutes or longer
7.3.1
Self-Test, Display, Key, and Buzzer
Press the
MENU
key. Select
1) MAINTENANCE
from the categories and
1) Self Test
from the param-
eters.
If an error occurs during the test, refer to Section 5.7, "Self Test" to verify the content of the error.
If any result of the following tests does not fall within the accuracy specifications, calibrate the 4601 as
outlined in Chapter 8, "CALIBRATION" or contact an ADC CORPORATION sales representative for the
calibration or servicing.