4601 I-V Meter Operation Manual
8.3 Calibration Points and Tolerance Ranges
8-7
Current measurement calibration
Current limit calibration
Current measurement calibration (VM/IM)
Current sink limit calibration (L limit)
Range
Calibration point
Tolerance
specification [digit]
Current limit
range
Calibration point
Tolerance
specification [digit]
300
A
000.000
A
2d
300
A
000.0
A
0.5d
-300.000
A
2d
-300.0
A
0.5d
3 mA
0.000 mA
0.5d
3 mA
0.00000 mA
2d
-3.000 mA
0.5d
-3.00000 mA
2d
30 mA
00.00 mA
0.5d
-30.00 mA
0.5d
30 mA
00.0000 mA
2d
300 mA
000.0 mA
0.5d
-30.0000 mA
2d
-300.0 mA
0.5d
3 A
0.000 A
0.5d
300 mA
000.000 mA
5d
-3.000 A
0.5d
-300.000 mA
5d
10 A
00.00 A
1d
-10.00 A
1d
3 A
0.00000 A
5d
-3.00000 A
5d
Current source limit (H limit)
10 A
00.0000 A
5d
Current limit
range
Calibration point
Tolerance
specification [digit]
-10.0000 A
5d
300
A
000.0
A
1.0d
+300.0
A
1.0d
3 mA
0.000 mA
1.0d
+3.000 mA
1.0d
30 mA
00.00 mA
1.0d
+30.00 mA
1.0d
300 mA
000.0 mA
1.0d
+100.0 mA
1.2d
3 A
0.000 A
1.0d
+0.100 A
1.2d
10 A
00.00 A
1.0d
+00.10 A
1.2d