4601 I-V Meter Operation Manual
1.1 Product Overview
1-2
Reference cell measurement terminal
•
Current measurement range:
-32.000 mA to +319.999 mA
•
Current range:
3mA, 30 mA, 300 mA
•
Current measurement resolution:
10 nA, 100 nA, 1
A
Voltage measurement terminal
•
Voltage measurement range:
3.19999 V
•
Voltage range:
30 mV, 300 mV, 3 V
•
Voltage measurement resolution:
0.1
V, 1
V, 10
V
Temperature measurement T-type thermocouple measurement terminal
•
Measurement range:
-50.00
C to +400.00
C
•
Measurement resolution:
0.01
C
Temperature measurement Pt measurement terminal
•
Measurement range:
Pt100
-200.00
C to +850.00
C
JPt100
-200.00
C to +649.00
C
•
Measurement resolution:
0.01
C
Temperature measurement AD590 measurement terminal
•
Measurement range:
-50.00
C to +150.00
C
•
Measurement resolution:
0.01
C
AD converter
Measurement terminal
Sweep source
mode
DC source
mode
ADC1
Voltage measurement (Vm)
ADC2
Current measurement (Im)
ADC3
Reference cell measurement terminal (Ir)
(fixed)
Select
Voltage measurement terminal (Em)
-
Select
T-type thermocouple measurement terminal (Tc1) Ch1
-
Select
T-type thermocouple measurement terminal (Tc2) Ch2
-
Select
Pt measurement terminal (Pt1) Ch1
-
Select
Pt measurement terminal (Pt2) Ch2
-
Select
IC sensor measurement terminal (Ad1) Ch1
-
Select
IC sensor measurement terminal (Ad2) Ch2
-
Select