4601 I-V Meter Operation Manual
3.2.1 Setting and Outputting Voltage Source Value
3-4
Example display
Press
again to stop the output (Standby). The
OPERATE
key goes OFF.
2. Setting the voltage source value using the rotary knob (for fixed range)
This section describes the procedure of setting the source value using the rotary knob and outputting
it.
Each procedure should be started with the initial state (factory default settings). If the 4601 is not in
the initial state, initialize the 4601 referring to Section 3.5, “Initializing Setting Conditions.” previ-
ously.
Output voltage of +10.000 V according to the following procedure:
Operating procedure
Check that the
FIT
key is OFF and the fixed range is selected.
Example display
Press
to switch to the 50 V range.
Example display
4
1
2
3
4
5 6
1
2