4601 I-V Meter Operation Manual
4.2.10 MENU Key (Parameter Setting)
4-8
F)
EXT.SIGNAL
Sets the external signal-related parameters.
1)
Operate Control
Selects the signal for the BNC connector
OPERATE IN/OUT
on the rear
panel.
Standby In
Sets Standby when the signal level changes from LO to HI.
OPR/STBY In
Sets Standby when the signal level changes from LO to HI.
Sets Operate when the signal level changes from HI to LO.
OPR/SUS In
Sets Suspend when the signal level changes from LO to HI.
Sets Operate when the signal level changes from HI to LO.
Operate Out
Outputs LO in the Operate status.
Outputs HI in the Standby or Suspend status.
2)
Comp./Sync. Sel
Selects the signal for the BNC connector
COMPLETE OUT/SYNC
OUT
on
the rear panel.
Sync. Out
Outputs the Sync. Out signal.
Complete Out
Outputs the Complete Out signal.
3)
Signal Width
Selects the signal width of Complete Out/Sync Out between 10
s and 100
s.
G)
PARAMETER
Loads or saves the setting parameters.
1)
Load
Loads the setting parameters saved in the non-volatile memory.
from User-0
Loads the parameters from the non-volatile memory,
User-0.
from User-1
Loads the parameters from the non-volatile memory,
User-1.
from User-2
Loads the parameters from the non-volatile memory,
User-2.
from User-3
Loads the parameters from the non-volatile memory,
User-3.
Default
Loads the factory default values as setting parameters.
2)
Save
Saves the setting parameters to the non-volatile memory.
to User-0
Saves the current parameters to the non-volatile memory,
User-0.
to User-1
Saves the current parameters to the non-volatile memory,
User-1.
to User-2
Saves the current parameters to the non-volatile memory,
User-2.
to User-3
Saves the current parameters to the non-volatile memory,
User-3.
Default to All
Saves the factory default values to the non-volatile
memory, User-0 to User-3.