4601 I-V Meter Operation Manual
5.2.2 Sweep Source Mode
5-9
The following table shows the source and measurement timings in the sweep source mode with reverse
OFF (one-way sweep).
Table 5-4 Source Measurement Operations in Sweep Source Mode
Operational
condition
Sampling
Description
Operation
Remarks
Operate ON
AUTO
Executes continu-
ous measurement
with the setting
period time Tp.
Th: Hold time
Tp: Period time
Td: Measurement
delay time
Tds:Source delay
time
Tm: Measurement
time
Integration
time + Mea-
surement data
processing
time
Tcn: Operate pro-
cessing time
Trd: Trigger delay
time
HOLD
Executes measure-
ment by trigger
input.
The step count per
trigger is 1.
Executes measure-
ment by trigger
input.
The step count per
trigger is other than
1.
• The bias value is output before sweep start.
• The start value is output when the start trigger is input.
• Sweep starts when the hold time has passed after start trigger input.
• As for external trigger input, the trigger is enabled when the trigger delay time has passed.
• When the sampling is AUTO, the sweep step changed with the period time.
• When the sampling is HOLD, the sweep step operates by every trigger input.