4601 I-V Meter Operation Manual
5.2.4 Output at Sweep End
5-11
5.2.4
Output at Sweep End
The output value at the sweep end can be switched by setting RTB.
Return to Bias
5.2.5
Source and Measurement Timings
5.2.5.1
Basic Source and Measurement Timings
The following figure shows the basic timings of voltage source and voltage and current measurements.
Figure 5-7 Basic Source and Measurement Timings
•
Set the measurement delay time Td longer than the settling time Ts to obtain stable measured values.
Ts < Td
RTB
Waveform
Description
ON
Returns to the bias
value at the sweep
end.
OFF
Holds the stop
value at the sweep
end.
Bias value
Start value
Stop value
Bias value
Start value
Stop value
Tp:
Measurement cycle in the DC source
mode
Step cycle in the sweep source mode
Tds:
Source delay time
Ts:
Settling time
Td:
Measurement delay time
Tm:
Measurement time
Tm=Tit+Tk
Tit:
Integration time
Tk:
Measurement data processing time