4601 I-V Meter Operation Manual
6.5 Data Output Format
6-15
6.5
Data Output Format
6.5.1
Measured Value Output
The 4601 outputs data in the order of
ADC1 measured value,
ADC2 measured value and
ADC3
measured value.
When data of multiple measurement memories is output, this order is repeated. (
...)
Whether each measured value is output or not can be set separately.
6.5.2
Output Target Selection
The measured values to be output can be set by using the OTM command or from the MENU.
*This setting is not initialized by RST or RINI.
:Output
-:No output
OTM
designated
value
ADC3
measured
value
ADC2
measured
value
ADC1
measured
value
1
-
-
ADC1 measured value only
2
-
-
ADC2 measured value only
3
-
ADC1 and ADC2 measured values
4
-
-
ADC3 measured value only
5
-
ADC1 and ADC3measured values
6
-
ADC2 and ADC3measured values
7
All measured values (ADC1 to ADC3)
1
2
3
1 2 3
1 2 3
ADC1 measured value
ADC2 measured value
ADC3 measured value
+
+
Internal data
ADC1 measured value
ADC2 measured value
ADC3 measured value
b7
b0