4601 I-V Meter Operation Manual
5.1.2 Reference Cell Measurement Terminal
5-2
Twist pair cables are connected to HI OUTPUT and LO OUTPUT and to HI SENSE and LO SENSE in
pairs respectively. The wire size is determined depending on the current value used on the OUTPUT side.
Table 5-1 below shows the allowable current value and its corresponding wire size.
Figure 5-2 Connection with Output Unit
5.1.2
Reference Cell Measurement Terminal
The following figure shows the terminal to measure the reference cell current.
Figure 5-3 Reference Cell Measurement Terminal
(A) Wiring Using Twisted Pair Cables
(B) Wiring Using Shielded Twisted Pair Cables
Table 5-1 Allowable Current Value
and Wire Size
Current value
Wire (AWG)
Up to 500 mA
24
Up to 2 A
22
Up to 3.2 A
18
Up to 5 A
16
Up to 10 A
14
Applicable plug:
Product equivalent to HIROSE ELECTRIC
HA316P-4S (71)
-E
+E
-I
+I