6240B DC Voltage Current Source/Monitor Operation Manual
6.5 Status Register Structure
6-11
The following figure shows the 6240B status register structure.
Figure 6-1 Status Register Structure
Status Register Structure
Device Event Status Register
Comparator HI
Comparator GO
Comparator LO
Arrive at Store Number
Suspend
Limiter Low
Limiter High
Ext. Operate Off In
Device Event Status Enable Register
Error Register
Measurement
data
Query data
Standard Event Status Register
*OPC Operation Complete
Power ON
Output buffer
Standard Event Status Enable Register
Status Byte Register
Service Request Enable Register
*Parameter Lost means save parameter data lost or backup parameter data lost.
Ext. Trigger In
Memory Full
Operate
Calibration End
Sweep End
Sweep Step Complete
End Of Measure
Self Test Error at Power ON
Self Test Error
Calibration Coefficient Lost
Over Load
Fan Stopped
Over Heat
Source Unit Error
Parameter Lost *1
Command Argument Error
Command Execution Error
Command Format Error
Unknown Command
Over Range
Calculation Erro