6240B DC Voltage Current Source/Monitor Operation Manual
9.1 Source and Measurement
9-5
Typical value of full scale item tolerance
The full-scale item tolerances listed below are added to the integration time 100
s to 10 ms and the 1-day
stability.
S/H:
Measurement in the sample hold mode (integration time: 100
s)
When LO OUTPUT is grounded to the chassis, the additional error of integration time in 30
A range is the
same as that in 300
A range.
Source linearity:
3 digits or less (
5 digits or less in the 4 A range)
Maximum output current:
0 to
15 V;
1 A (DC)
0 to
10 V;
4 A (maximum pulse width 20 ms/duty factor
20 %
Maximum compliance voltage: Up to 1 A (DC);
0 to
15 V
Up to 4 A (pulse); 0 to
10 V
Voltage source range
Voltage limiter range
A
Rm
B
Is
Tolerance
30 mA
300 mV
10
V
10 m
3
A
10 mA
1 m
300 mA
300 mV
10
V
30
A
100 mA
103
1 A
300 mV
10
V
300
A
1 A
13
30 mA
300 mV
10
V
100 m
3
A
10 mA
1 m
300 mA
300 mV
10
V
30
A
100 mA
130
1 A
300 mV
10
V
300
A
1 A
40
30 mA
300 mV
10
V
1
3
A
10 mA
1.3 m
300 mA
300 mV
10
V
30
A
100 mA
400
1 A
3 V
50
V
300
A
1 A
350
Measurement
range
Integration time Unit: digits (at 5
1
/
2
digit display)
10 ms
5 ms
1 ms
500
s
100
s
S/H
Voltage
measurement
300 mV
10
15
20
30
60
200
3 V
5
8
10
15
30
50
15 V
5
8
10
15
30
50
Current
measurement
30
A
200
300
300
300
300
300
300
A
20
30
30
30
70
100
3 mA
10
30
30
30
50
80
30 mA
10
30
30
30
50
100
300 mA
10
15
15
15
50
100
1 A
20
30
75
75
250
500
4 A
20
30
75
75
250
500