6240B DC Voltage Current Source/Monitor Operation Manual
9.2 Source and Measurement Function
9-9
9.2
Source and Measurement Function
DC source and measurement:
Source and measurement of DC voltage and current
Pulse source and measurement:
Source and measurement of pulse voltage and current
(However, measurement auto range in pulse source is impossible)
DC sweep source and measurement: Source and measurement by linear, 2- slope linear, random and fixed
levels
Pulse sweep source and measurement:
Source and measurement by linear, 2- slope linear, random and fixed
levels (However, measurement auto range in pulse source is impossible)
Low-resistance measurement:
Measurement by pulse current source voltage measurement
Integration time:
9 types: 100
s, 500
s, 1 ms, 5 ms, 10 ms, 1 PLC, 100 ms, 200 ms and
S/H
S/H: Sample hold (integration time: 100
s) measurement
(Enabled only in the pulse source or pulse sweep source modes)
(PLC: Power Line Cycle 50 Hz: 20 ms, 60 Hz: 16.66 ms)
Sweep mode:
Reverse ON (round) / OFF (one way)
Sweep repeat count:
1 to 1000 times or infinite
Maximum number of sweep steps:
8000 steps
Maximum random sweep memory: 8000 data
Measurement data memory:
8000 data
Measurement auto range:
Available only in VSIM or ISVM
Measurement function link mode:
Links the source function to the measurement function.
(VSIM or ISVM)
ON/OFF available
Limiter:
The HI and LO limit values can be set individually.
(However, setting the current limit values with the same polarity is not
allowed.)
Calculation function:
NULL calculation
Comparator calculation (HI, GO, or LO)
Scaling calculation
MAX, MIN, AVE and TOTAL calculations
Trigger style:
Auto trigger, External trigger
Output terminal:
Front;
Safety socket
HI OUTPUT, HI SENSE, LO OUTPUT, and LO SENSE
Maximum input:
15 V peak (between HI-LO)
2 V peak (between OUTPUT and SENSE)
250 V maximum (between LO and chassis)
Maximum remote sensing voltage:
1 V Max; HI OUTPUT - HI SENSE, LO OUTPUT - LO SENSE
(The voltage between HI SENSE and LO SENSE must be within the
maximum output voltage range.)