6240B DC Voltage Current Source/Monitor Operation Manual
8.5 Calibration Procedure
8-7
Figure 8-4 Calibration Procedure (2)
Voltage source
/limiter calibration
VS mode
[Calibration mode setting]
[Range setting]
Enters the DMM data input mode.
VS-Hi mode
VS-Lo mode
DMM data
acquisition
Within
tolerance range?
XD DMM data
DMM data input *Out-of-range setting results in a parameter error.
XD DMM data
DMM data input *Out-of-range setting results in a parameter error.
DMM data
acquisition
Checks the calibration result.
DMM data
acquisition
Calibration result data
save and display
DMM data
acquisition
Calibration result data
save and display