7351 Series Digital Multimeter Operation Manual
6.7.1 Selecting the Command Language
6-17
6.7.1.1
Compatibility
If the R6451 command is selected, this instrument receives the R6451 command and executes it.
CAUTION:
1.
In the compatibility mode, note that the operations such as the measurement timing may be different from
other commands.
2.
Commands, which are related to non-existent functions in this instrument such as the 4-20 mA measurement
or memory card control, cannot be used.
By selecting the R6451 language, commands for R6441 can be received and executed.
CAUTION: When using commands for R6441, ensure the following:
1.
For the DC voltage measurement function, the R2 (20 mV range) command for R6441 sets R3 (200 mV range)
for this instrument. The measured value is output with the 200 mV range format.
2.
For the DC current measurement function, the R5 (20 mA range) command for R6441 sets R6 (200 mA range)
for this instrument. The measured value is output with the 200 mA range format.
3.
The F14 (High-speed AC voltage measurement function) command cannot be used. Substitute the F2 (AC
voltage measurement function) command.
4.
The F34 (High-speed AC current measurement function) command cannot be used. Substitute the F6 (AC
current measurement function) command.
5.
A mantissa of the output data format is one digit more than R6441.
Command enabled when R6451 is selected Command disabled when R6451 is selected
E
M n
Z
BZ n
BZ?
MS
IDN?
CS
SB? (RS-232 only)
TRSn
MSR?
MSEn
MSE?
QSR?
QSEn
QSE?
*PSC
*PSC?
*ESR?
*ESE
*ESE?
LOP
LOP?
MIP
MIP?
HIP
HIP?
BP
BP?
OSE
OSE?