7351 Series Digital Multimeter Operation Manual
6.7.3 ADC Command Reference
6-26
Calculation dB/dBm calculation
DB0
dB calculation OFF
DB1
dB calculation ON
For the voltage and current measurement functions, ON
can be set.
DB2
dBm calculation ON
For the voltage measurement function, ON can be set.
DB?
Reply: One of DB0 to DB2
-
KDn
Constant D
n:0.000001E-9 to 999999.E+6
(1)
KDM
Sets Constant D to the measurement value.
KD?
Reply: KD
d.dddddE
dd *2
MAX•MIN
calculation
:
Header (Refer to
Section 6.7.2,
“Output Data
Format.”)
MN0
MAX•MIN calculation OFF
MN1
MAX•MIN calculation ON
*3
MN?
Reply: One of MN0 to MN1
-
MAX?
MAX value reading *2
Reply: M
d.dddddE
dd
-
MIN?
MIN value reading *2
Reply: I
d.dddddE
dd
-
AVE?
AVE value reading*2
Reply: A
d.dddddE
dd
-
AVN?
Measurement count reading *2
Reply: AVN
d.dddddE
dd
-
Comparator
calculation
CO0
OFF
CO1
ON
CO?
Reply: CO0 or CO1
-
HIn
HIGH constant
n:-999999.E+6 to +999999.E+6
Setting resolution: 0.000001E-9
(0)
LOn
LOW constant
n:-999999.E+6 to +999999.E+6
Setting resolution:0.000001E-9
(0)
HIM
Sets the HIGH constant to the measurement value.
LOM
Sets the LOW constant to the measurement value.
HI?
Reply: HI
d.dddddE
dd *2
-
LO?
Reply: LO
d.dddddE
dd *2
-
*2:
The position of the decimal point in the reply is fixed.
*3:
MN1/MN2 (MAX calculation ON/MIN calculation ON)
Item
Command
Description
Initial
value