7351 Series Digital Multimeter Operation Manual
6.7.3 ADC Command Reference
6-27
Calculation
PASS condition
range setting
LOP0
Calculation result LO is not specified as the PASS condi-
tion.
LOP1
Calculation result LO is specified as the PASS condition.
LOP?
Reply: LOP0 or LOP1
MIP0
Calculation result GO is not specified as the PASS condi-
tion.
MIP1
Calculation result GO is specified as the PASS condition.
MIP?
Reply: MIP0 or MIP1
HIP0
Calculation result HI is not specified as the PASS condi-
tion.
HIP1
Calculation result HI is specified as the PASS condition.
HIP?
Reply: HIP0 or HIP1
Statistical
SIRDn,m Statistical calculation range setting and performing the
calculation
n, m: 0 to 19999
Note:
1. An error occurs if no data exists in the range set
by the measurement data memory.
2. No parameters can be omitted. (0, -1)
(0, -1)
-
SIRD?
Statistical calculation range reading
Reply:SIRDddddd, ddddd
Initial state: SIRD+00000, -00001
-
SCNT?
Number of samples reading
Reply: SCNT+d.dd *2
-
SMAX? Maximum value reading
Reply: SMAX
d.dddddE
dd *2
-
SMIN?
Minimum value reading
Reply: SMIN
d.dddddE
dd *2
-
SAVE?
Average value reading
Reply: SAVE
d.dddddE
dd *2
-
SSIG?
Standard deviation value reading
Reply: SSIG
d.dddddE
dd *2
However, if the number of samples is 1 or less:
Outputs SSIG+9.30
-
SPTP?
MAX-MIN reading
Reply: SPTP
d.dddddE
dd *2
-
*2:
The position of the decimal point in the reply is fixed.
Item
Command
Description
Initial
value