7351 Series Digital Multimeter Operation Manual
7.1.7 Diode Measurement
7-9
7.1.7
Diode Measurement
The specifications other than the above are the same as those of 2000
range in 2W
measurement.
7.1.8
Continuity Measurement
The specifications other than the above are the same as those of 2000
range in 2W
measurement.
Range
Maximum display
Resolution
Measurement
current
Measurement accuracy *1
(% of r digits)
Temperature
coefficient
(ppm of r
digits)/
C
FAST
MED/
SLOW1, 2
FAST
MED/
SLOW1, 2
FAST
MED
SLOW1, 2
AZ ON AZ OFF
2000 mV
1999.9
1999.99 100
V
10
V
1 mA 0.014+2
0.014+5
0.014+3
15+0.25
15+1.5
*1
The offset error, which is calculated by the following formula; (the measurement cable resi 0.3
)
1 mA, is
added to the measurement accuracy
For AZ ON. For AZ OFF, two digits are added to the digit section.
Range
Maximum display
Resolution
Measurement
current
Measurement accuracy *1
(% of r digits)
Temperature
coefficient
(ppm of r
digits)/
C
FAST
MED/
SLOW1, 2
FAST
MED/
SLOW1, 2
FAST
MED
SLOW1, 2
AZ ON AZ OFF
2000
1999.9
1999.99 100 m
10 m
1 mA 0.014+2
0.014+5
0.014+3
15+0.25
15+1.5
*1
For AZ ON. For AZ OFF, two digits are added to the digit section.