7351 Series Digital Multimeter Operation Manual
8.3.3 Calibration Point
8-7
LP-2W
200
Short
190
-
-
2000
Short
1900
-
-
20 k
Short
19 k
-
-
200 k
Short
190 k
-
-
2000 k
Short
1900 k
-
-
20 M
Short
19 M
-
-
DCI
200 mA
Open
190 mA
-
-
2000 mA
Open
1900 mA
-
-
10 A
Open
10 A
-
-
ACI
ACI(AC+DC)
200 mA
Open
5 mA, 1 kHz
19 mA, 1 kHz
190 mA, 1 kHz
2000 mA
Open
50 mA, 1 kHz
190 mA, 1 kHz
1900 mA, 1 kHz
10 A
Open
100 mA, 1 kHz
1 A, 1 kHz
10 A, 1 kHz
Table 8-1 Calibration Point (2 of 2)
Measurement
function
Range
Zero
Calibration point1 Calibration point2 Calibration point3