7351 Series Digital Multimeter Operation Manual
4.2.1 Measurement Functions
4-9
Figure 4-6 AC Voltage Measurement (ACV and ACV (AC+DC))
Measurement method
TrueRMS
Input range
5% or more of a full scale
Crest factor
3 (This is restricted to the maximum allowable applied voltage.)
Filter response time
Approx. 1 sec (Time until the measurement value reaches within 0.1%
of the final value)
4.2.1.3
Resistance Measurement (2W
and LP-2W
)
1. Connect a DUT to the
and COM terminals on the front panel.
2. Press .
The resistance is measured by using the 2W
.
3. Press
and
.
The resistance is measured in low-power in which the measurement current is reduced.
WARNING:
Do not apply voltage that exceeds the maximum allowable applied voltage. Fire or electric
shocks due to the failure of this instrument may occur.
Table 4-4 Maximum Allowable Applied Voltage (2W
and LP-2W
)
Between
and COM terminals
1000 Vpeak
Between COM terminal and chassis (grounding)
500 V
SHIFT