7351 Series Digital Multimeter Operation Manual
7.3 General Specifications
7-15
7.3
General Specifications
Item
Specifications
Operation environment
Ambient temperature:0
C to +50
C
Relative humidity:
85%RH or less no condensation
Storage environment
Ambient temperature:-25
C to +70
C
Warm-up time
60 minutes or more
Display
7-digit fluorescent display A digit is displayed by seven segments.
Range switching
Automatic and manual
Input method
Floating
Measurement method
Integration
Overload display
OL
Trigger function
External trigger (7351A, 7351E + Option 03)
Memory
Data memory: Up to 20,000 data
Condition setting memory: 4 (USER0 to USER3)
Power supply
AC power supply:
100 V/120 V/220 V/240 V
(Can be switched by user)
Specify the option when ordering.
Use a power cable and fuse that are compliant with the safety standard when
changing the power supply voltage.
Power supply frequency
50 Hz/60 Hz
Power consumption
20 VA or below
Dimensions
Approx. 212 (width)
88 (height)
340 (depth) mm
Mass
3.4 kg or less
Safety
Compliance with IEC61010-1 Ed.2, Category II
EMI
Compliance with EN61326 classA
Compliance with FCC classA
Option number
Standard
OPT.32
OPT.42
OPT.44
Power supply voltage
100 V
120 V
220 V
240 V