7351 Series Digital Multimeter Operation Manual
8.3.4 Calibration Procedure
8-8
8.3.4
Calibration Procedure
Perform calibration in following procedure.
1. Set the calibration mode.
2. Short-circuit the V
and COM terminals and then open the mA and 10 A termi-
nals.
3. Calibrate the zero point in a function and range to be calibrated.
4. Connect the calibration standard to the input terminal.
5. Calibrate the full-scale point in a function and range to be calibrated.
6. Cancel the calibration mode.
Setting the calibration mode
1. Select ON from ‘CAL’ in
in
.
2. Press
to determine the setting.
The CAL indicator is turned on and indicates that the calibration mode is set.
3. Press
to exit from MENU.
4. Set a function to be calibrated.
Calibrating the zero point
5. Short-circuit between the V
and COM terminals and open the mA and 10 A ter-
minals.
6. Set a range to be calibrated.