7351 Series Digital Multimeter Operation Manual
5.3.2 Auto-range
5-9
5.3.2
Auto-range
When Auto-range is set to ON, the measurement range of this instrument is automatically set to the opti-
mum range depending on the input voltage.
The judgment value used when the range is changed
Range DOWN
When the displayed value is less than 10% of the displayed range (less than 200 V if
the DCV 1000 V range and 700 V range are both selected)
Range UP
When the displayed value exceeds the maximum display
Auto-range is set to OFF under the following conditions.
•
Frequency measurement
•
Diode measurement
•
Continuity measurement
•
The 10 A range for the DC and AC current measurements