7352 Series Digital Multimeter Operation Manual
6.6.4 SCPI Command Reference
6-36
Range configuration and settable maximum value
Measurement
condition
Temperature
sensor
[:SENSe]:TEMPerature:TCouple:
TYPE {K|T}
K: K-type thermocouple
T: T-type thermocouple
[:SENSe]:TEMPerature:TCouple:
TYPE?
Reply: K or T
Ach
Bch
Measurement
range notation
DCV
ACV
ACV
(AC+DC)
2W
DCI
ACI
ACI
(AC+DC)
LP-
2W
Freq
DCV
DCI
ACI
ACI
(AC+DC)
2000 nA
+1.99999E-06
20
A
+1.99999E-05
200
A
200
A
200
A
+1.99999E-04
2000
A
2000
A
2000
A
+1.99999E-03
20 mA
20 mA
20 mA
+1.99999E-02
200 mV
200 mV
200 mV
200 mA
200 mA
200 mA
(200 mV)
200 mV
+1.99999E-01
2000 mV 2000 mV
2000 mV
2000 mA 2000 mA
2000 mA
(2000 mV) 2000 mV
+1.99999E-00
-
-
-
-
-
10 A 10 A
10 A +1.01
20 V
20 V
20 V
(20V)
20 V
+1.01
200 V
200 V
200 V
200
200
(200V)
200 V
+1.02
-
700 V
700 V
-
-
(700V)
+7.02
1000 V
-
-
+1.03
2000
2000
+1.03
20 k
20 k
+1.04
200 k
200 k
+1.05
2000 k
2000 k
+1.06
20 M
20 M
+1.07
200 M
+1.08
NOTE: The range setting is not allowed for the continuity, diode, and temperature measurements. If the range setting is executed, an error
occurs.
Item
SCPI command
Description
Initial