7352 Series Digital Multimeter Operation Manual
6.6.5 Status Register Structure
6-47
A description of each bit in the status byte register is shown in Table 6-4.
Conditions which clear the status byte register
• When the power is turned on.
• When the *CLS command is executed. However, the MAV bit is not cleared if data exists in the
output buffer.
• The status byte register is not cleared by executing *STB? command.
Conditions which clear the service request enable register
• When the power is turned on. (When the PSC flag is set to 1)
• When the *SRE0 command is executed.
Table 6-4 Status Byte Register (STB)
bit
Name
Description
0
MSB
Measurement
Summary Bit
ON: When any event of Measurement Event Register occurs and Measurement
Event Register is set to 1, this bit is set to 1 if the corresponding bit of the
Measurement Event Enable Register is set to 1.
OFF: This bit is set to 0 when Measurement Event Register is cleared by being
read.
1
Not used
Always 0
2
EAV
Error Available
ON: This bit is set to 1 when the error information is stored in Error Queue.
OFF: This bit is set to 0 when the error information is read from Error Queue and
no data has existed in Error Queue.
3
QSB
Questionable
Summary Bit
ON: When any event of Questionable Event Register occurs and Questionable
Event Register is set to 1, this bit is set to 1 if the corresponding bit of the
Questionable Event Enable Register is set to 1.
OFF: This bit is set to 0 when Questionable Event Register is cleared by being
read.
4
MAV
Message Available
ON: This bit is set to 1 when the output data is entered in the output buffer.
OFF: This bit is set to 0 when the data is read from the output buffer.
5
ESB
Standard Event Status
ON: When any event of SESR occurs and SESR is set to 1, this bit is set to 1 if
the corresponding bit of the SESER is set to 1.
OFF: This bit is set to 0 when SESR is cleared by being read (*ESR?).
6
MSS
Master Summary
ON: When any event of STB occurs, this bit is set to 1 if the corresponding bit of
the SRER is set to 1.
RQS
Request Service
ON: If MSS is set to 1, SRQ is generated and RQS is set to 1.
OFF: When STB is read in the serial polling, RQS is set to 0.
7
OSB
Operation Summary
Bit
ON: When any event of Operation Event Register occurs and Operation Event
Register is set to 1, this bit is set to 1 if the corresponding bit of the Opera-
tion Event Enable Register is set to 1.
OFF: This bit is set to 0 when Operation Event Register is cleared by being read.