7352 Series Digital Multimeter Operation Manual
7.1.2 AC Voltage Measurement
7-3
7.1.2
AC Voltage Measurement
7.1.2.1
Ach AC Voltage Measurement (ACV-Ach and ACV(AC+DC)-Ach)
Measurement method
True RMS measurement, RMS display
Input range
5% or more of a full scale
Crest factor
3:1 at a full scale (This is restricted to the maximum allowable applied
voltage.)
Temperature coefficient
(1/10 of measurement accuracy that includes the additional error)/
C in each
range and frequency range
Response time
Approx. 1 s (Time until the measurement value reaches within 0.1% of the
final value in the same range)
ACV-Ach
Input
terminal
Range
Maximum display
Resolution
Input impedance
FAST/MED
SLOW1, 2
FAST/MED
SLOW1, 2
V
Hz-
COM A
200 mV
199.99
199.999
10
V
1
V
1 M
2%, 140 pF or less
2000 mV
1999.9
1999.99
100
V
10
V
20 V
19.999
19.9999
1 mV
100
V
200 V
199.99
199.999
10 mV
1 mV
700 V
749.9
749.99
100 mV
10 mV
Range
Measurement accuracy *1
(% of r digits)
20 Hz -
45 Hz
45 Hz -
100 Hz
100 Hz -
20 kHz
20 kHz -
50 kHz
50 kHz -
100 kHz
200 mV
0.38+140
0.11+120
0.1+100
0.25+150
0.7+240
2000 mV
0.38+140
0.11+120
0.1+100
0.2+150
0.6+240
20 V
0.38+140
0.11+120
0.1+100
0.2+150
0.6+240
200 V
0.38+140
0.11+120
0.1+100
0.2+150
0.6+240
700 V
0.38+100
0.11+100
0.1+100
–
–
*1:
For a sine-wave input When RATE is set to FAST or MED, the digit term is reduced
to 1/10 of its specification.