7352 Series Digital Multimeter Operation Manual
7.1.3 Ach Resistance Measurement (2W
-Ach and LP-2W
-Ach)
7-5
7.1.3
Ach Resistance Measurement (2W
-Ach and LP-2W
-Ach)
2W
resistance measurement
Low power resistance measurement (LP-2W
)
Response time:
200 M
= 2 s
(Time until the measurement value reaches within 0.1% of the final value)
20 M
= 0.5 s
(Time until the measurement value reaches within 0.1% of the final value)
Voltage between input terminals without connections:
7.5 V or less
Maximum allowable applied voltage
Input
terminal
Range
Maximum display
Resolution
Measurement
current
Measurement accuracy *1
(% of r digits)
Temperature
coefficient
(ppm of reading
+ digits)/
C
FAST
MED/
SLOW1, 2
FAST
MED/
SLOW1, 2
FAST
MED
SLOW1, 2
Auto-
zero
ON
Auto-
zero
OFF
V
Hz-
COM A
200
199.99
199.999
10 m
1 m
1 mA
0.02+4
0.02+9
0.02+8
20+1
20+2
2000
1999.9
1999.99 100 m
10 m
1 mA
0.02+2
0.02+5
0.014+3
15+0.25
15+1.5
20 k
19.999
19.9999
1
100 m
100
A
0.02+2
0.02+5
0.014+3
15+0.25
15+1.5
200 k
199.99
199.999
10
1
10
A
0.02+2
0.02+5
0.02+5
20+0.25
20+1.5
2000 k
1999.9
1999.99
100
10
1
A
0.03+2
0.03+10
0.03+10
35+2
35+5
20 M
19.999
19.9999
1 k
100
100 nA
0.2+2
0.2+10
0.2+10
155+2
155+5
200 M
199.99
199.999
10 k
1 k
10 nA
1.5+2
1.5+10
1.5+10
1500+2
1500+5
Input
terminal
Range
Maximum display
Resolution
Measurement
current
Measurement accuracy *1
(% of r digits)
Temperature
coefficient
(ppm of reading
+ digits)/
C
FAST
MED/
SLOW1, 2
FAST
MED/
SLOW1, 2
FAST
MED
SLOW1, 2
Auto-
zero
ON
Auto-
zero
OFF
V
Hz-
COM A
200
199.99
199.999
10 m
1 m
1 mA
0.02+4
0.02+9
0.02+8
20+1
20+5
2000
1999.9
1999.99 100 m
10 m
100
A
0.03+4
0.03+9
0.03+8
20+1
20+3
20 k
19.999
19.9999
1
100 m
10
A
0.03+4
0.03+9
0.03+8
20+1
20+3
200 k
199.99
199.999
10
1
1
A
0.03+4
0.03+9
0.03+8
30+1
30+3
2000 k
1999.9
1999.99
100
10
100 nA
0.2+4
0.2+12
0.2+12
150+2
150+5
20 M
19.999
19.9999
1 k
100
10 nA
1.5+5
1.5+50
1.5+50
1500+2
1500+5
*1:
For Auto-zero ON For Auto-zero OFF, 2 digits are added to the digit term. In addition, the resistance of the measurement cable and
an offset error of 0.3
are added.
Between V
Hz and COM A terminals
1000 Vpeak
Between COM A terminal and the chassis
500 V
Between COM A and COM B terminals
200 V