7352 Series Digital Multimeter Operation Manual
4.2.1 Measurement Functions
4-11
Figure 4-7 Resistance Measurement (2W
-Ach and LP-2W
-Ach)
The measurement result includes the input cable resistance and contact resistance. To reject these re-
sistance effects, short the tip of the input cable and set the NULL calculation to ON before the mea-
surement.
Ensure the following when in the high-resistance measurement to reduce the effect of the induction
noise.
•
Shielding the DUT.
•
Using the A01001 input cable (sold separately) whose HI side is shielded.
•
Securing the input cable to prevent movement.
Table 4-4 Maximum Allowable Applied Voltage (2W
-Ach, LP-2W
-Ach)
Input terminal
Maximum allowable applied voltage
Ach
Between V
Hz and COM A terminals
1000 Vpeak
Between COM A terminal and the chassis
500 V
Between COM A and COM B terminals
200 V