4.2.1 Measurement Functions
4-15
4.2.1.7
Ach Diode Measurement
1. Connect a DUT to the Ach V and COM terminals on the front panel.
2. Press
and .
The diode indicator (
) is activated.
A current of approximately 1 mA is applied from the COM terminal to the V terminal, and the voltage
between both terminals is measured.
Measurement range
2000 mV
Measurement current
1 mA
WARNING:
Do not apply voltage that exceeds the maximum allowable applied voltage. Fire or electrical
shock due to the failure of this instrument may occur.
Figure 4-11 Diode Measurement-Ach
Table 4-8 Maximum Allowable Applied Voltage (Diode-Ach)
Input terminal
Maximum allowable applied voltage
Ach
Between V
Hz and COM A terminals
1000 Vpeak
Between COM A terminal and the chassis
500 V
Between COM A and COM B terminals
200 V
SHIFT