7352 Series Digital Multimeter Operation Manual
4.2.3 Changing the Measurement Speed and Display Digit
4-25
•
Dual sequential measurement
Measurement when the same channel function of Ach or Bch is set for the right and left displays
The measurement time is expressed as the following formula:
Measurement time for the left d Time for function change (left to right) + Measurement time
for the right d Time for function change (right to left) + Data processing time
Table 4-18 Measurement Speed during the HOLD ON (Trigger source: BUS) RS-232 (7352A)
•
Single measurement
Measurement only for the left display
•
Dual synchronous measurement
Measurement when the Ach function is set for the left display
and the Bch function is set for the right display
Sampling
rate
Measurement speed (Measurement time)
Integration
time
Sampling
rate
Measurement speed (Measurement time) *1
Integration
time
DCV-Ach
DCV-Bch
DCI-Ach
DCI-Bch
2W
-Ach
LP-2W
-Ach
ACV-Ach
ACI-Ach
ACI-Bch
Continuity
-Ach
Diode-Ach
ACV(AC+DC)
-Ach
ACI(AC+DC)
-Ach
ACI(AC+DC)
-Bch
TEMP
FREQ
-Ach
Ach
Bch
DCV-Ach
DCI-Ach
2W
-Ach
LP-2W
-Ach
ACV-Ach
ACI-Ach
Continuity
-Ach
Diode-Ach
DCV-Bch
DCI-Bch
ACI-Bch
FAST
19 times/s (51 ms)
18 times/s
(53 ms)
38 times/s
(26 ms)
2 ms
FAST
18.5 times/s (54 ms)
2 ms
MED
(50 Hz)
14 times/s (69 ms)
11 times/s
(90 ms)
8.6 times/s
(116 ms)
1 PLC
MED
(50 Hz)
13.9 times/s (68 ms)
1 PLC
MED
(60 Hz)
15 times/s (65 ms)
11 times/s
(85 ms)
8.6 times/s
(116 ms)
1 PLC
MED
(60 Hz)
14.7 times/s (68 ms)
1 PLC
SLOW1
6 times/s (150 ms)
3 times/s
(251 ms)
0.9 times/s
(1016 ms)
100 ms
SLOW1
6 times/s (153 ms)
100 ms
SLOW2
4 times/s (248 ms)
2 times/s
(449 ms)
-
200 ms
SLOW2
4 times/s (251 ms)
200 ms
Conditions: Auto-range: OFF, Auto-zero: OFF, Calculation: OFF, Display:
OFF, and Trigger delay: 0 sec
Conditions: Auto-range: OFF, Auto-zero: OFF, Calculation:
OFF, and Display: OFF
*1:
For the dual synchronous measurement, if ACV(AC+DC)-Ach, ACI(AC+DC)-Ach, ACI(AC+DC)-Bch, TEMP, or FREQ-Ach is
set for the Ach or Bch function, the measurement time is the same as the single measurement time for the set function.