7352 Series Digital Multimeter Operation Manual
7.1.4 DC Current Measurement (DCI)
7-6
7.1.4
DC Current Measurement (DCI)
7.1.4.1
Ach DC Current Measurement (DCI-Ach)
Input
terminal
Range
Maximum display
Resolution
Resistance
between
terminals
(A fuse
resistance is
included.)
Measurement accuracy *1
(% of r digits)
Temperature
coefficient
(ppm of reading
+ digits)/
C
FAST
MED/
SLOW1, 2
FAST
MED/
SLOW1, 2
FAST
MED
SLOW1, 2
Auto-
zero
ON
Auto-
zero
OFF
mA-
COM A
2000 nA
1999.9
1999.99
100 pA
10 pA
11.5 k
or
less
0.15+35 0.15+40
0.15+40
150+4
150+4
20
A
19.999
19.9999
1 nA
100 pA
0.1+7
0.1+7
0.1+5
100+1
100+1.6
200
A
199.99
199.999
10 nA
1 nA 102
or less
0.04+20 0.04+40
0.04+40
40+4
40+5
2000
A
1999.9
1999.99
100 nA
10 nA
0.04+2
0.04+7
0.04+5
40+1
40+1.6
20 mA
19.999
19.9999
1
A
100 nA
2
or less
0.04+20 0.04+40
0.04+40
40+4
40+5
200 mA
199.99
199.999
10
A
1
A
0.05+2
0.05+7
0.05+6
50+0.6
50+3
2000 mA
1999.9
1999.99
100
A
10
A
0.6
or less
0.15+2
0.15+7
0.15+6
150+0.6
150+3
*1:
For Auto-zero ON For Auto-zero OFF, 2 digits are added to the digit term.
Maximum allowable
applied current
Input protection
Fuse replacement
Between mA and
COM A terminals
2000 mA 2 A/250 V fast-blow fuse which is com-
pliant with IEC60127 sheet1
On the rear panel