7352 Series Digital Multimeter Operation Manual
7.1.7 Temperature Measurement (TEMP)
7-13
7.1.7
Temperature Measurement (TEMP)
Maximum allowable applied voltage
Compliant thermocouple standard:
JIS C1602
Cold junction compensation:
Internal
7.1.8
Ach Diode Measurement-Ach
Other specifications are the same as the resistance measurement function.
Thermocouple
K(CA)
T(CC)
Input terminal
Range
-50
C to 1370
C
-50
C to 400
C
TEMP HI - LO
Maximum display
13700
4000
Resolution
0.1
C
0.1
C
Measurement accuracy
0.15%
3
C
0.15%
3
C
Between TEMP (HI/LO) and COM A terminals
200 Vpeak
Between TEMP (HI/LO) and COM B terminals
200 Vpeak
Between TEMP (HI) and TEMP (LO) terminals
36 Vpeak
Between TEMP (HI/LO) terminal and the chassis
500 V
Input
terminal
Range
Maximum display
Resolution
Measurement
current
Measurement accuracy *1
(% of r digits)
Temperature
coefficient
(ppm of reading
+ digits)/
C
FAST
MED/
SLOW1, 2
FAST
MED/
SLOW1, 2
FAST
MED
SLOW1, 2
Auto-
zero
ON
Auto-
zero
OFF
V
Hz-
COM A
2000 mV
1999.9
1999.99
100
V
10
V
1 mA 0.014+2 0.014+5
0.014+3
15+0.25
15+1.5
*1:
For Auto-zero ON For Auto-zero OFF, 2 digits are added to the digit term. In addition, an offset error, which is obtained by multiply-
ing (the resistance of the measurement cable + 0.3
) by 1 mA, is added.