7352 Series Digital Multimeter Operation Manual
4.2.1 Measurement Functions
4-21
Measurement method
TrueRMS
Input range
5% or more of a full scale
Crest factor
3 (However, this value is limited by the current protection fuse rating.)
Filter response time
Approx. 1 sec (Time until the measurement value reaches to within
0.1% of the final value)
Restrictions on the auto-range
The 10 A range is a fixed range.
4.2.1.12
Temperature Measurement
Connect a thermocouple to the TEMP HI and LO terminals on the rear panel.
Press .
The ‘
C’ indicator is activated.
Temperatures at the terminals are internally compensated.
For the thermocouple, select K or T type from ‘TEMP’ in the
category of
.
Figure 4-16 Temperature Measurement
Table 4-14 Maximum Allowable Applied Voltage (TEMP)
Input terminal
Maximum allowable applied voltage
TEMP
Between TEMP HI and TEMP LO terminals
36 Vpeak
Between TEMP HI/LO terminals and the chassis
500 V
Between TEMP HI/LO terminals and COM A terminal
200 V
Between TEMP HI/LO terminals and COM B terminal
200 V
TEMP
MEAS
MENU
Thermocouple (K or T)