7352 Series Digital Multimeter Operation Manual
5.7.2 Dual Sequential Measurement
5-15
Figure 5-2 Single Measurement Sequence
5.7.2
Dual Sequential Measurement
Measurement is performed when functions of the same channel, Ach or Bch, are set for the right and left
measurement displays.
For more information on combinations of related functions, see Table 5-4.
When one trigger event is received, measurements are performed according to the following order:
Figure 5-3 shows the dual sequential measurement sequence.
The following shows an operation example to set the dual sequential measurement during single measure-
ment.
Key operation
Display section
1
The left display shows the state of the DCV function.
DCV-Ach
measured value
2
Press .
DCV-Ach
measured value
2nd
3
Press
. The set function is displayed on the right
side display section.
In this case, the DCI-Ach function is displayed.
DCV-Ach
measured value
DCI-Ach
measured value
Trigger
Measurement
Calculation
Data output
Td (Trigger delay time)
Td (Trigger delay time)
Measurement
1
Measurement
2
Calculation 1
Calculation 2
Data 1
Data 2
(a) HOLD on / Trigger
(b) HOLD off (Free run)
Measurement
Calculation
Data output
Measurement
1
Measurement
2
Measurement
3
Measurement
4
Measurement
5
Calculation 4
Calculation 3
Calculation 2
Calculation 1
Data 1
Data 2
Data 3
Data 4
Trigger
Measurement for the left side display
Changing the function and range
Measurement for the right side display
Data output
2nd
DCI