7352 Series Digital Multimeter Operation Manual
5.9 Calculation Functions
5-24
Figure 5-7 Calculation System Diagram
NOTE:
1.
The following results may be obtained from calculating the data. This is because the internal resolutions of mea-
surement values and setting values for the calculation are higher than the resolutions of the displayed data and
remote output data.
• If the buzzer is set to ON in the MAX/MIN calculation, the buzzer sounds when the MAX or MIN value is
renewed. However the buzzer may sound even when the display and remote output data are not changed.
2.
If the calculation is set to ON, the display of the calculation result is different from the ordinary measurement
values and the decimal point and unit do not depend on the current measurement range.
<Example>
DC voltage measurement, measurement range: 2000 mV, input voltage: 1 V
Measurement value: 1000.00 mV
Calculation Result: 1.00000 V
3.
If a calculation is performed while the ‘2nd’ indicator is on, it is applied to a measurement for the right display.
For the comparator calculation and the calculation between 2 measurements, however, it is not applied.
Measurement
data on the
left side
display
NULL
Smoothing
Scaling
Comparator
dB
dBm
Data
memory
Statistics
Calculation result
Maximum value
Minimum value
Average value
Maximum value
Minimum value
Average value
Standard deviation
Dispersion
Left side display
Measurement
data on the
right side
display
NULL
Smoothing
Scaling
dB
dBm
Right side display
Calculation
between 2
measurements
Data
memory
Statistics
Calculation result
Maximum value
Minimum value
Average value
Maximum value
Minimum value
Average value
Standard deviation
Dispersion