7352 Series Digital Multimeter Operation Manual
5.9.3 Scaling Calculation
5-28
5.9.3
Scaling Calculation
1. Function
The scaling calculation equation is shown below.
When SCL is selected from the MATH item in
, the SCL indicator in the display section
turns on and the scaling calculation is performed.
If OFF is selected from the MATH item in
while the scaling calculation is performed, the
calculation is canceled and the SCL indicator turns off.
CAUTION:
The actual measured value is not displayed while the scaling calculation is performed.
Remember that any hazardous voltage present on the input connector or test lead may not
detected at this time.
2. Setting scaling constants
NOTE:
Press the
key if quitting the setting procedure.
1. Set Constants A, B, and C in
.
2. After selecting a constant to be set, the constant can be set by pressing
.
3. Setting the numerical value of the constant
The setting ranges of Constants A, B, and C are shown below.
Constant
Setting range
Minimum setting
value
A
-999999.E +6 to +999999.E + 6
0.00001E - 9
B
0.00000E - 9
C
The exponential part is set by using a unit prefix (n,
, m, k, M).
Display =
Constant C
Measured value M - Constant B
Constant A
MENU
MENU
MENU
CALC
MATH
OFF/dB/dBm/SCL
EXIT/LOCAL
MENU
VAL A
VAL B
VAL C
MENU
CONST