7451A/7461A/7461P Digital Multimeter Operation Manual
7.1.7 Diode Measurement
7-15
Gate time
Maximum allowable applied voltage
7.1.7
Diode Measurement
•
7461A/61P
•
7451A
Sampling rate
Gate time
Measurement
frequency range
Maximum
measurement
period
Maximum display
7461A/
61P
7451A
SLOW
1000 ms
1 Hz to 300 kHz
2.2 s
9999999
999999
MED
100 ms
10 Hz to 300 kHz
220 ms
999999
99999
FAST
10 ms 100 Hz to 300 kHz
22 ms
99999
9999
Between V and COM terminals
700 Vrms, 1000 Vpeak, 2.2
10
7
V•Hz
Between COM terminal and chassis
500 V
Range
Maximum
display
Resolution
Measurement
current
1000 mV
1199.999
1
V
1 mA
The specifications other than the above are the same as those of 1000
range in 2W
measurement.
However, the offset error, which is calculated by the following formula; (the measurement cable resi internal circuit
resistance error of 0.2
)
1 mA, is added to the measurement accuracy in 2W
measurement.
Range
Maximum
display
Resolution
Measurement
current
3000 mV
3199.99
10
V
1 mA
The specifications other than the above are the same as those of 3000
range in 2W
measurement.
However, the offset error, which is calculated by the following formula; (the measurement cable resi internal circuit
resistance error of 0.2
)
1 mA, is added to the measurement accuracy in 2W
measurement.