Mounting and connection
MSX-ilog-xx-xx
www.addi-data.com
28
ADDI-DATA
®
PARTNER FÜR PRÄZISION
Display
Meaning
Possible cause
Recommendation
Lights red
The data logger
cannot start up.
Hardware error
(e.g. RAM)
Contact us for a replacement or
repair.
Flashes red
Harmful ambient
conditions
The internal
temperature is
outside the working
range (< -40 °C or
> +85 °C).
- The data logger should be
quickly exposed to warmer or
colder temperatures (according
to the working range
temperatures).
- Under these conditions, the
measurement values may be
inaccurate and the whole
operation of the data logger be
restricted.
- Conditions of this kind may
damage the internal
components and therefore
make the whole data logger
unusable.
Flashes red/green
The data logger is
working and
communicating
correctly, but
possible risks may
be anticipated.
The internal
temperature is LOW
or HIGH.
- The data logger should be
quickly exposed to warmer or
colder temperatures (according
to the working range
temperatures).
Flashes
red/yellow
The data logger
cannot be accessed
via a network
connection, and
possible risks may
be anticipated.
The internal
temperature is LOW
or HIGH
and
the
network cables are
not connected.
- The data logger should be
quickly exposed to warmer or
colder temperatures (according
to the working range
temperatures).
- Check the Ethernet connection
(see LEDs “Port 0 ACT/LINK”
and “Port 1 ACT/LINK”).
- Note that other faults may also
occur in the meantime, which
prevent data acquisition
(e.g. a short-circuit).