11
CMRR: >100 dB (50 or 60 Hz)
Note 3:
The excitation power supply for RTD measure is 0.2mA, and each position has 4-wire, 3-wire and 2-wire methods. The accuracy
is as follows:
(1) The accuracy data given in the table is based on the 4-wire mode;
(2) The accuracy of the 3-
wire method is 10mΩ more than the accuracy of the 4
-wire mode;
(3) The accuracy of the 2-
wire method is 50mΩ more than the accuracy of the 4
-wire mode.
Note 4:
Minimum threshold voltage for frequency and pulse measure: 2.5V
Note 5:
Supported frequency units: Hz, kHz, MHz, CPM, CPH, s, ms, us;
Note 6
: Optional rising edge or descending edge trigger mode for pulse measurement.
1.3.3 Electrical Output Specifications
Table 4 Electrical Output Specifications
Specifications
Range
Resolution
Accuracy
ADT226
ADT227/ ADT227-HART
Voltage DC
150 to 150 mV
5 μV
0.015% RDG + 15
uV
0.005% RDG + 15
μV
-1.5 to 1.5 V
0.05 mV
0.015%RDG + 0.15mV
0.005% RDG + 0.15 mV
-15 to 15 V
0.5 mV
0.015%RDG + 1.5 mV
0.005% RDG + 1.5 mV
Current DC
0 to 25 mA
0.5 uA
0.015% RDG + 1.2 uA
0.01% RDG +
1.2 μA
Resistance
0 to 400 Ω
10
mΩ
0.015% RDG +
20 mΩ
0.005% RDG +
20 mΩ
0 to 4000 Ω
10
0 mΩ
0.015% RDG +
200 mΩ
0.01% RDG +
200 mΩ
Frequency
(0.01 ~ 5) Hz
0.00001 Hz
0.005% RDG + 0.00005 Hz
0.002%RDG + 0.00002 Hz
Summary of Contents for 226
Page 1: ...226 227 Multifunction Process Calibrator...
Page 2: ......
Page 3: ...ADT226 227 Multifunction Process Calibrator User Manual Version 2306V01 Additel Corporation...
Page 4: ......
Page 6: ......
Page 31: ...19 1 4 Basic Structure Figure 1 Basic Structure...
Page 33: ...21 Figure 2 Power adaptor...
Page 64: ...52 External module information including external module A and external module B...
Page 107: ...95...